A secondary standard for ambient dose equivalent, H*(10), is necessary for the dissemination of the unit Sievert (Sv), but there is no such standard commercially available currently. Furthermore, the measurement of H*(10) instead of calculating H*(10) from air kerma and conversion coefficients is needed for unknown radiation fields. We developed a prototype of a new secondary standard for H*(10) based on a spherical 1l ionization chamber for air kerma. This chamber was modified with copper wires at the inner surface to adjust the response of the chamber according to H*(10). Additionally, a Makrolon shell and an aluminium coating were added to optimize the response at energies below 50keV. The prototype fulfils the requirements given in ISO 4037-2 in the energy range from 12keV to 7MeV. In combination with an electrometer, it can be used as area dosemeter, suitable for pulsed fields and for low energy radiation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.