Abstract
Objective To develop a new silicone coating iodine-125(s-125I) radioactive stent, and to compare it with the traditional titanium coating iodine-125(t-125I) radioactive stent. Methods Medical silicone solution was mixed with 125I solution, evenly applied to specific locations of stents to make the silicone-coated 125I stents. The total radioactivity of 125I solution was 24mCi, by which 20 silicone-coated 125I stents were made. A total of 20 traditional titanium stents of the same radioactivity were collected, and the utilization rate of the radionuclides of the two types of stents were compared. Results With the same amount of radioactivity, the s-125I stent showed a higher utilization rate of nuclide[91.2%(21 878/24 000)VS 50.1%(12 031/24 000)](P<0.05). Conclusion The s-125I stent has a high utilization rate of nuclide, which can save the cost and energy largely. However, the application of the s-125I stent still needs further research. Key words: Bile duct neoplasms; Iodine-125; Silica gel; Stents
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.