Abstract
We developed a new non-derivatization analytical method for the determination of galactose in the diagnosis of galactosemia by high-performance anion-exchange chromatography (HPAEC)-pulsed amperometric detection (PAD). With an anion-exchange column, the analytes were separated efficiently using 3 mM NaOH containing 1 mM NaOAc, and 200 mM NaOH was added for post-column reagent. The limit of detection (S/N = 3) and limit of quantification (S/N = 10) for galactose were 25 ng/mL and 83 ng/mL, respectively. Linear dynamic range was from 4.67 mg/dL to 53.46 mg/dL ( r 2 = 0.9999). The mean recovery of galactose for intra-, inter-day assays were found to be of satisfactory results (98.14–101.42%).
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.