Abstract

A new simple, accurate, economic, rapid and precise reverse phase high performance liquid chromatographic method has been developed for the validated of Paracetamol and Metoclopramide HCl, in its pure form as well as in pharmaceutical dosage form. Chromatography was carried out on X bridge C18 (4.6×150mm) 5 µ column using a mixture of Methanol: Phosphate Buffer pH-3.6 (30:70v/v) as the mobile phase at a flow rate of 1.0ml/min, the detection was carried out at 260nm. The retention time of the Paracetamol and Metoclopramide HCl was 2.669, 3.855±0.02min respectively. The method produce linear responses in the concentration range of 10-50µg/ml of Paracetamol and 10-50µg/ml of Metoclopramide HCl. The method precision for the determination of assay was below 2.0%RSD. The method is useful in the quality control of bulk and pharmaceutical formulations.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call