Abstract

A turbidimetric-kinetic photometer which monitors 64 samples simultaneously and independently has been developed for the automated measurement of endotoxin. This instrument measures the turbidity change at 660 nm in a gelation reaction of the LAL/endotoxin system and automatically scores the reaction time required to obtain a certain turbidity change as gelation time. Since a stationary dry bath incubator controls the reaction temperature of the samples at 37 ± 0.5°C and also controls the ambience of the optical system assembled into the incubator, the objective detection of gelation is provided without any disturbance from sample vibration, and the drift of the detection system is minimized. An extremely wide range for the calibration curve, such as 4 or 5 orders of magnitude of endotoxin concentration, is obtained without sample dilution. The coefficients of variation of the calculated endotoxin concentration are 4.91 to 14.0%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.