Abstract

A capacitive probe [Tan et al., Rev. Sci. Instrum. 88, 023502 (2017)] is one of a few diagnostics that is directly sensitive to the plasma potential. Using this diagnostic technique, a Multi-channel Linear Capacitive Probe (MLCP) is developed for turbulence measurements. The MLCP has 10 spatial channels and provides 9 points of radial electric field measurements simultaneously with a spatial step of 7 mm. A new readout circuit and a correction technique for low frequency attenuation are also developed to achieve the required spatial and time resolution. A performance test of the MLCP using a reversed field pinch plasma confirms that the MLCP resolves sub-centimeter structures of the equilibrium radial electric field profile and fluctuations up to 680 kHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.