Abstract
An extrapolation chamber of type Böhm 23392, used for the dosimetry of RQR beam qualities at the Secondary Standard Dosimetry Laboratory of King Faisal Specialist Hospital, is presented. A computer-based motorized system consisting of a stepping motor coupled to the chamber's built-in micrometer screw was designed to expedite the measurements process, giving a linear relationship between the number of motor steps and the chamber depth with a Coefficient of Determination (COD) equal to 0.9990. The extrapolation chamber along with the motorized system was used to determine the extrapolation curves. The slope of the extrapolation curves, measured for the eight RQR beam qualities with a relative uncertainty between 0.17% and 0.58%, varies linearly with the beam quality expressed in terms of HVL, with a COD between 0.9995 and 1.0000.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.