Abstract
In this study, a robust, selective and simplified method was developed and validated for the simultaneous quantitative analysis of 23 underivatized amino acids in human serum using mixed-mode chromatography with tandem mass spectrometry (LC–MS/MS). Serum samples were deproteinized with acetonitrile and subjected to LC–MS/MS analysis. The chromatographic separation of amino acids was achieved using a mixed-mode column (150×3mm, 3μm) with a gradient elution system; the mobile phase consisted of 50mM ammonium formate and 0.1% formic acid in acetonitrile. The total run time was 22min. Eluted compounds were detected in the electrospray ionization-positive mode with multiple reaction monitoring. The validation study evaluated linearity, repeatability, intra and inter-day accuracy and precision, and matrix effect. The validation results were satisfactory in all the tested parameters. This method was successfully applied to the analysis of amino acids in the clinical sample of human serum.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.