Abstract
We fabricated a transition-edge sensor (TES) microcalorimeter for spectroscopic measurements of L X-rays emitted by transuranium (TRU) elements. A Pu L X-ray energy spectrum was obtained by irradiation with L X-rays emitted by a 244Cm source to the TES microcalorimeter. The energy resolution was evaluated to be 64 eV at the full width at half maximum (FWHM) for L X-rays at 14.28 keV. The obtained L X-ray spectrum indicated that the TES microcalorimeter made it possible to distinguish between the L X-ray lines of TRU elements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.