Abstract
A new fiber-optic pH sensor system has been developed. The sensor uses an absorbtive indicator compound with a long wave-length absorption peak near 625 nm; change in absorption over the pH range 6.8 to 7.8 is reasonably linear. The sensor is interrogated by a pulsed, red LED. Return light signal is split into short and long wave-length components with a dichroic mirror; the respective signals are detected by photodiodes, and their photocurrents are used to form a ratiometric output signal. In laboratory tests, the sensor system provided resolution of 0.01 pH, accuracy of +/- 0.01 pH, and response time of 30-40 s. Following gamma sterilization, laboratory sensor testing with heparinized human blood yielded excellent agreement (e.g., r = 0.992 for n = 42) with a clinical blood gas analyzer. Excellent sensor performance and low cost, solid-state instrumentation are hallmarks of this sensor-system design.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.