Abstract
A highly accurate (+/-3%) mechanical loading and measurement system combined with a trabecular bone diffusion culture-loading chamber has been developed, which provides the ability to study trabecular bone (and possibly) cartilage under controlled culture and loading conditions over long periods of time. The loading device has been designed to work in two main modes, either to apply a specific compressive strain to a trabecular bone cylinder or to apply a specific force and measure the resulting deformation. Presently, precisely machined bone cylinders can be loaded at frequencies between 0.1 Hz to 50 Hz and amplitudes over 7,000 microepsilon. The system allows accurate measurement of many mechanical properties of the tissue in real time, including visco-elastic properties. This paper describes the technical components, reproducibility, precision, and the calibration procedures of the loading system. Data on long term culture and mechanical responses to different loading patterns will be published separately.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.