Abstract
Using manganese dioxide as a solid-phase oxidant, a post-column HPLC reactor for paralytic shellfish poison toxins was constructed and evaluated. Operating parameters such as reaction temperature and pH, flow-rate, reactor column size, and MnO 2 particle size were studied. Based on a 3:1 signal-to-noise ratio, the limits of detection for the non-hydroxylated toxins were in the sub-nanogram range, while the hydroxylated toxins were ten times less detectable. The limit of detection for saxitoxin was about 0.1 ng per injection. The repeatability of replicate injections was ≤10% (relative standard deviation). The system was used to analyze extracts of shellfish and plankton, yielding results that were in agreement with those obtained by established analytical methods. Extracts of shellfish contaminated at the regulatory limit of 0.8 μg/g were successfully analyzed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.