Abstract

In this study, a simple and rapid sample preparation method named liquid-nitrogen-induced homogeneous liquid-liquid microextraction has been developed for the extraction and pre-concentration of Co(II) and Ni(II) ions before their analysis by flame atomic absorption spectrometry. For this purpose, first, acetonitrile containing 8-hydroxyquinoline is added into a sample solution and the mixture is vortexed. As a result, a homogeneous solution is formed. Subsequently, the solution is cooled using liquid nitrogen for a few seconds. By this process, due to difference in the freezing point of acetonitrile and water, the homogeneous state is broken and the analytes (as oxinate complexes) are extracted into liquid acetonitrile phase collected on top of the frozen aqueous phase. The linear dynamic ranges obtained for Ni(II) and Co(II) were 1.0-30 and 0.50-20μgL-1, respectively. The obtained limits of detection were 0.36 and 0.20μgL-1 for Ni(II) and Co(II), respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.