Abstract

A reversed-phase high performance liquid chromatography (RP-HPLC) method, with acetic acid and sodium perchlorate phase modifiers, was developed to separate a mixture of linear alkylbenzene sulphonates (LAS), containing 10 to 13 carbon atoms. The effects of methanol-water compositions and concentrations of used modifiers were investigated and compared. The separation achieved with 50 mg L−1 acetic acid was found satisfactory whereas a concentration of 10 g L−1 sodium perchlorate was preferred. Chromatograms obtained with UV and Fourier transform infrared (FTIR) detection showed almost similar features and HPLC-FTIR interface spectra of LAS components exhibited excellent agreement of absorption features to those of standard FTIR spectrum and no thermal degradation was found to occur.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call