Abstract

A specific and sensitive homologous radioimmunoassay has been developed for secreted hamster prolactin. Hamster serum and pituitary extracts showed parallel dilution-response curves with hamster prolactin. The sensitivity of the assay ranged from 0.5 to 1.0 ng/ml, and the intra- and inter-assay coefficients of variation were 6 and 10%, respectively. Additionally we have demonstrated that the rat prolactin radioimmunoassay kit distributed by the National Institute of Arthritis, Metabolism, and Digestive Diseases is an inadequate method for the measurement of hamster prolactin.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.