Abstract

A new glass capillary plate (CP) has been developed for a hole-type micro-pattern gas detector (MPGD). The developed CP has a funnel structure (funnel-CP) fabricated using a special manufacturing process for a CP. The opening to surface area ratio is 83% for the funnel-CP. A higher ratio is expected to improve the collection efficiency of primary electrons created by X-ray radiations, a higher optical transparency should prove reliable for digital X-ray radiography and cold neutron imaging. Basic performance tests of the hole-type MPGD were carried out using X-ray beams for several gas mixtures based on neon and argon at 1atm. We successfully obtained a gain of up to 1×104, and an energy resolution of 16% for 6keV X-rays. An excellent imaging capability was demonstrated by the X-ray image.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.