Abstract

Determination of cholic acid is very important and necessary in body fluids due to its both pharmaceutical and clinical significance. In this study, a quartz crystal microbalance (QCM) nanosensor, which is imprinted cholic acid, has been developed for the assignation of cholic acid. The cholic acid selective memories have been generated on QCM electrode surface by using molecularly imprinted polymer (MIP) based on methacryloylamidohistidine-copper (II) (MAH-Cu(II)) pre-organized monomer. The cholic acid imprinted nanosensor was characterized by atomic force microscopy (AFM) and then analytical performance of the cholic acid imprinted QCM nanosensor was studied. The detection limit was found to be 0.0065μM with linear range of 0.01–1000μM. Moreover, the high value of Langmuir constant (b) (7.3*105) obtained by Langmuir graph showed that the cholic acid imprinted nanosensor had quite strong binding sites affinity. At the last step of this procedure, cholic acid levels in body fluids were determined by the prepared imprinted QCM nanosensor.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.