Abstract

Abstract A very sensitive, highly selective and reversible optical chemical sensor (optode) for mercury ion is described. The sensor is based on the interaction of Hg2+ with 2-mercapto-2-thiazoline (MTZ) in plasticized PVC membrane incorporating a proton-selective chromoionophore (ETH5294) and lipophilic anionic sites (sodium tetraphenylborate, NaTPB). The membranes were cast onto glass substrates and used for the determination of mercury ion in aqueous solutions by batch and flow-through methods. The sensor could be used in the range 2.0 × 10−10 to 1.5 × 10−5 M (0.04 ng mL−1 to 3 μg mL−1) Hg2+ with a detection limit of 5.0 × 10−11 M and a response time of

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.