Microwave digestion and isotope dilution inductively coupled plasma mass spectrometry (ID-ICP-SFMS) has been applied to the determination of Pb in rice flour. In order to achieve highly precise determination of low concentrations of Pb, the digestion blank for Pb was reduced to 0.21 ng g(-1) after optimization of the digestion conditions, in which 20 mL analysis solution was obtained after digestion of 0.5 g rice flour. The observed value of Pb in a non-fat milk powder certified reference material (CRM), NIST SRM 1549, was 16.8 +/- 0.8 ng g(-1) (mean +/- expanded uncertainty, k = 2; n = 5), which agreed with the certified value of 19 +/- 3 ng g(-1) and indicated the effectiveness of the method. Analytical results for Pb in three brown rice flour CRMs, NIST SRM 1568a, NIES CRM 10-a, and NIES CRM 10-b, were 7.32 +/- 0.24 ng g(-1) (n = 5), 1010 +/- 10 ng g(-1) (n = 5), and 1250 +/- 20 ng g(-1) (n = 5), respectively. The concentration of Pb in a candidate white rice flour reference material (RM) sample prepared by the National Metrology Institute of Japan (NMIJ) was observed to be 4.36 +/- 0.28 ng g(-1) (n = 10 bottles).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.