A new type of high sensitivity counter employing a liquid scintillation cell as a coincidence gate for a 3″ NaI(Tl) well crystal has been developed for the detection of small amounts of X-γ emitting nuclides. This counter has approximately 10 times the efficiency of a comparable detector employing a conventional gas proportional counter as the coincidence gate. The background of the existing system under the full width at half maximum of the 54Mn photopeak is 0.02 cpm. Measurements to the 15% precision level of as little as 0.25 dpm of 54Mn in lunar rock samples are reported. The applicability of the counter to X-γ emitters other than 54Mn is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.