Abstract

The purpose of this study is to develop a geometrically accurate imaging protocol at 3 T magnetic resonance imaging (MRI) for stereotactic radiosurgery (SRS) treatment planning. In order to achieve this purpose, a methodology is developed to investigate the geometric accuracy and stability of 3 T MRI for SRS in phantom and patient evaluations. Forty patients were enrolled on a prospective clinical trial. After frame placement prior to SRS, each patient underwent 3 T MRI after 1.5 T MRI and CT. MR imaging protocols included a T1-weighted gradient echo sequence and a T2-weighted spin echo sequence. Phantom imaging was performed on 3 T prior to patient imaging using the same set-up and imaging protocols. Geometric accuracy in patients and phantoms yielded comparable results for external fiducial reference deviations and internal landmarks between 3 T and 1.5 T MRI (mean ⩽0.6 mm; standard deviation ⩽0.3 mm). Mean stereotactic reference deviations between phantoms and patients correlated well (T1: R = 0.79; T2: R = 0.84). Statistical process control analysis on phantom QA data demonstrated the stability of our SRS imaging protocols, where the geometric accuracy of the 3 T SRS imaging protocol is operating within the appropriate tolerance. Our data provide evidence supporting the spatial validity of 3 T MRI for targeting SRS under imaging conditions investigated. We have developed a systematic approach to achieve confidence on the geometric integrity of a given imaging system/technique for clinical integration in SRS application.