Abstract

We have recently started to develop a high-resolution gamma-ray spectrometer for material defect analysis. Our gamma-ray detector is a microcalorimeter consisting of an iridium/gold bilayer transition edge sensor (TES) thermometer and a bulk Pb absorber, which is directly coupled to a TES with Stycast 2850FT epoxy. This paper describes our TES based gamma-ray detector and the first experimental results for 662 keV and 1173 keV incident gamma-rays. We report an energy resolution of 4.7 keV at 662 keV and 2.9 keV at 1173 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.