Abstract

Purpose:The purposes of this study were two-fold: first, to develop a four-axis moving phantom for patient-specific quality assurance (QA) in surrogate signal-based dynamic tumor-tracking intensity-modulated radiotherapy (DTT-IMRT), and second, to evaluate the accuracy of the moving phantom and perform patient-specific dosimetric QA of the surrogate signal-based DTT-IMRT.Methods:The four-axis moving phantom comprised three orthogonal linear actuators for target motion and a fourth one for surrogate motion. The positional accuracy was verified using four laser displacement gauges under static conditions (±40 mm displacements along each axis) and moving conditions [eight regular sinusoidal and fourth-power-of-sinusoidal patterns with peak-to-peak motion ranges (H) of 10–80 mm and a breathing period (T) of 4 s, and three irregular respiratory patterns with H of 1.4–2.5 mm in the left–right, 7.7–11.6 mm in the superior-inferior, and 3.1–4.2 mm in the anterior–posterior directions for the target motion, and 4.8–14.5 mm in the anterior–posterior direction for the surrogate motion, and T of 3.9–4.9 s]. Furthermore, perpendicularity, defined as the vector angle between any two axes, was measured using an optical measurement system. The reproducibility of the uncertainties in DTT-IMRT was then evaluated. Respiratory motions from 20 patients acquired in advance were reproduced and compared three-dimensionally with the originals. Furthermore, patient-specific dosimetric QAs of DTT-IMRT were performed for ten pancreatic cancer patients. The doses delivered to Gafchromic films under tracking and moving conditions were compared with those delivered under static conditions without dose normalization.Results:Positional errors of the moving phantom under static and moving conditions were within 0.05 mm. The perpendicularity of the moving phantom was within 0.2° of 90°. The differences in prediction errors between the original and reproduced respiratory motions were −0.1 ± 0.1 mm for the lateral direction, −0.1 ± 0.2 mm for the superior-inferior direction, and −0.1 ± 0.1 mm for the anterior–posterior direction. The dosimetric accuracy showed significant improvements, of 92.9% ± 4.0% with tracking versus 69.8% ± 7.4% without tracking, in the passing rates of γ with the criterion of 3%/1 mm (p < 0.001). Although the dosimetric accuracy of IMRT without tracking showed a significant negative correlation with the 3D motion range of the target (r = − 0.59, p < 0.05), there was no significant correlation for DTT-IMRT (r = 0.03, p = 0.464).Conclusions:The developed four-axis moving phantom had sufficient accuracy to reproduce patient respiratory motions, allowing patient-specific QA of the surrogate signal-based DTT-IMRT under realistic conditions. Although IMRT without tracking decreased the dosimetric accuracy as the target motion increased, the DTT-IMRT achieved high dosimetric accuracy.

Highlights

  • Several respiratory motion management techniques, including forced shallow breathing, breath-holding, respiratory gating, and dynamic tumor tracking (DTT), have been proposed to reduce the uncertainty caused by respiratory motion,2 thereby reducing the probability of normal tissue complications

  • We evaluated the accuracy of this phantom and performed patient-specific dosimetric quality assurance (QA) of surrogate signal-based DTT-Intensity-modulated radiotherapy (IMRT)

  • We developed a four-axis moving phantom that reproduced patient 3D tumor and 1D surrogate motions to realize truly patient-specific dosimetric QA for surrogate signal-based DTT-IMRT

Read more

Summary

Introduction

Intensity-modulated radiotherapy (IMRT) has been used in clinical practice for a variety of tumor types and anatomical locations. when treating tumors, those located in the thoracic and abdominal regions, respiratoryinduced tumor motion leads to uncertainty during beam delivery. Compared with three-dimensional (3D) conformal radiotherapy, a complex interplay occurs between the motion of the multileaf collimator (MLC) and that of the tumor in IMRT, leading to underdosage or overdosage in portions of the target volume and/or surrounding tissues.2–6Several respiratory motion management techniques, including forced shallow breathing, breath-holding, respiratory gating, and dynamic tumor tracking (DTT), have been proposed to reduce the uncertainty caused by respiratory motion, thereby reducing the probability of normal tissue complications. Of these, recent interest has focused on the DTT technique, which is able to dynamically reposition the radiation beam or the robotic couch in accordance with the position of the target. DTT decreases internal uncertainties due to target motion, without the need for a prolonged treatment time or the burden of shallow breathing or breath-holding for patients. Compared with three-dimensional (3D) conformal radiotherapy, a complex interplay occurs between the motion of the multileaf collimator (MLC) and that of the tumor in IMRT, leading to underdosage or overdosage in portions of the target volume and/or surrounding tissues.. Several respiratory motion management techniques, including forced shallow breathing, breath-holding, respiratory gating, and dynamic tumor tracking (DTT), have been proposed to reduce the uncertainty caused by respiratory motion, thereby reducing the probability of normal tissue complications.. Recent interest has focused on the DTT technique, which is able to dynamically reposition the radiation beam or the robotic couch in accordance with the position of the target.. DTT decreases internal uncertainties due to target motion, without the need for a prolonged treatment time or the burden of shallow breathing or breath-holding for patients. There are two approaches to achieving DTT: direct and indirect tracking methods.. There are several approaches to localize the internal target position: dual fluoroscopic imaging systems, an electromagnetic transponder, fourdimensional (4D) computed tomography (CT), combined 4DCT and daily cone-beam CT (CBCT) projection images, and 4DCBCT.

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.