Abstract

We developed a timing counter with a monolithic array of multi-pixel photon counters (MPPCs) coupled to a <formula formulatype="inline" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"><tex Notation="TeX">$10 \times 12.6 \times 24~\hbox{mm}^3$</tex></formula> plastic scintillator. A signal of the MPPC was processed and its fast components were amplified by a developed readout circuit to improve the timing. The performances of the counter were evaluated by electrons with energies of 855 MeV. A set of the MPPC based scintillation counter and a PMT based reference counter achieved the TOF resolution less than 80 ps ( <formula formulatype="inline" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"><tex Notation="TeX">$\sigma $</tex></formula> ). The gain was almost constant up to the count rate of 170 kHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.