Abstract

A disposable prototype pyruvate biosensor was constructed using pyruvate oxidase immobilised on mediated meldolas blue electrodes to determine pungency in onions (Allium cepa L.). The optimum operating potential was +150 mV (versus Ag/AgCl). A strong correlation between the biosensor response and untreated onion juice of known pyruvate concentration 2-12 micromol/g fresh weight (FW) was demonstrated. The biosensor was able to differentiate between low and high pungency onions. The detection limit using 1 unit of pyruvate oxidase was 1-2 micromol/g FW. Optimum concentrations of co-factors TPP, FAD and MgSO4 comprising the enzyme cocktail were determined as being 0.04, 0.1 and 30 mM, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call