Abstract

A disposable ion-selective optode for mercury based on trityl-picolinamide (T-Pico) as neutral ionophore was developed. The sensing layer consist of plasticised PVC incorporating T-Pico as a selective ionophore for Hg 2+, ETH 5418 as a chromoionophore, and potassium tetrakis[3,5-bis(trifluoromethyl)phenyl] borate as lipophilic salt. The measurement principle is based on an ion-exchange mechanism. When the optode membrane is introduced into a water sample for 5 min, there is a colour change from red to blue, depending on the mercury concentration (pH 4.7), making it possible to use the absorbance at 665 nm as the analytical signal. The optode membrane response to Hg 2+ was not fully reversible; however, it reveals a very good selectivity to many cations including alkali, alkaline earth, transition, and heavy metal ions. The detection limit for Hg 2+ is 5.0 × 10 −7 M at pH 4.7. The response characteristics of the sensor including dynamic range, reproducibility, response time, and lifetime are discussed in detail. This sensor was used for the determination of mercury in environmental water samples with satisfactory recovery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.