Abstract
A continuous radon (Rn-222) concentration monitoring system for use in underground soil was developed and tested. The system consists of a 19-mm-diameter, 1100-mm-long detector assembly and a microprocessor based data logger. A small volume chamber is installed at the tip of the detector assembly. A thin ZnS(Ag) scintillator film inside the chamber and a photomultiplier tube (PMT) detect alpha particles from radon and its daughters. When the system is in measurement, the detector part is buried into underground soil. An energy resolution of approximately 70% full width half maximum (FWHM) was obtained for 5.5 MeV alpha particles from Am-241. Both the rise time and fall time for the system were measured to be approximately 1-2 h. Temporal variations in underground radon concentration at different depths were investigated simultaneously using four sets of the developed system. The results confirmed that the developed system is useful for continuous measurement of radon concentration in underground soil.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.