Abstract

A flow-through cell of the confined wall-jet type for electrochemical surface plasmon resonance (ESPR) applications has been developed. The geometrical cell volume of the electrochemical (EC) detection is adjusted to the surface plasmon resonance (SPR) cell volume by reducing the electrode diameter. An electrode diameter of 3 mm gives the best similarity of the SPR and EC responses, using hexacyanoferrate(II) in a flow-injection analysis (FIA) setup. Further hydrodynamic characterization is done by continuous-flow analysis. The dependence of the limiting Faradaic current on the volume flow rate ( φ v) and the spacing ( b) between working and auxiliary electrode have been determined. The exponents found for φ v are in the range from 0.29 to 0.35 and for b in the range from −0.37 to −0.57 for typical flows (10–1000 μl min −1) and spacings (0.13–0.50 mm), respectively. Preliminary results of an impedimetric/liposome-enhanced SPR immunosensor for interferon-γ (INF-γ), based on an acetylcysteine self-assembled monolayer, show the characteristics and applicability of this new flow-through cell for ESPR research in a FIA setup.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.