Abstract

The present study aimed to develop a simple computer simulation method of low-dose radiographs based on a radiograph acquired at a clinical-dose level. A chest phantom was used for the development of this method. In this method, a simulated low-dose image was obtained from a clinical-dose image using an input-output characteristic curve of a flat panel detector and noise metrics of the standard deviation (SD) and noise power spectrum. We applied this method for low-dose images of a chest phantom to evaluate the simulation accuracy. The noise SDs were compared between the simulated and real images corresponding to 1/2, 1/4, and 1/8 of clinical doses. The relative error of noise SDs in the chest phantom images was less than 3%. Therefore, we believe that the proposed simulation method has the potential to be useful for determination of the optimal exposure condition in chest radiography to reduce patients' exposure dose.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.