Abstract

SUMMARYThis research aims to develop a compact synchrotron dedicated for proton beam therapy using a scanning irradiation method. The effective length and magnetic uniformity of the bending magnets in the synchrotron were analyzed by 3D static magnetic field calculations. The calculation results indicate that the shape of the bending magnet satisfies the specification for beam stability. A push–pull multifeed‐driven technique allows shortening the length of the FINEMET RF acceleration cavity for the synchrotron from 600 mm to 450 mm. The circumference of the synchrotron is 18 m, which is the world's most compact size for proton beam therapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.