Abstract
A scanned proton beam yields dose distributions that in most cases aresuperior to passively scattered proton beams and to other external radiationtreatment modalities. The present paper gives a description of the scanningsystem that has been developed at the Svedberg Laboratory (TSL) in Uppsala.The scanning technique and the technical design are described. The solutionwith a small pole gap of the magnets and a moveable second magnet results in avery compact scanning head, which can therefore be incorporated in a gantry ofrelatively limited size. A prototype was constructed that has been used torealize various dose distributions with a scanned beam of 180 MeV protons atTSL.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.