Abstract

A new neutron reflectometer, CN REF-V, has been installed in the cold-neutron laboratory building at the HANARO, a research reactor in Korea. The instrument has a vertical scattering plane and uses a constant wavelength of λ = 4.7535 A, and it is monochromated by using pyrolytic graphite (PG) crystals. Its measurable minimum reflectivity and maximum momentum transfer for typical solid films are 10-8 and 0.3 A-1, respectively. A liquid-nitrogen-cooled beryllium filter is used to remove λ/2 contamination due to the PG (002) crystals. With a gold wire activation analysis method, the neutron flux at its sample position was measured and found to be 5.67 × 105 neutrons/cm2 /s. Reflectivity measurements of thin films were successfully carried out with the instrument. A detailed characterization of the instrument and the results of the reflectivity measurements are described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.