Abstract
In this work, an analytical flow system able to perform low pressure chromatography with amperometric detection is presented. As case study, the determination of niacin (vitamin B3) in coffee brewed samples was selected. The manifold comprised a 1.0cm length monolithic column coated with didecyldimethylammonium bromide, a laboratory-made boron doped diamond electrode, and featured in-line ionic strength adjustment of the mobile phase. The figures of merit concerning the selected case study namely, detection limit, 7.90×10−7M, determination rate, ca. 10 samplesh−1, mobile phase and ISA solution consumption, ca. 2.6mL per analysis, and CV, below 5% for retention time and peak height, showed the competitiveness of this analytical strategy comparing to the described HPLC methods for niacin determination. The strategy displays a simple configuration, low cost, fast and easy assembling, foreseeing its use to general purpose applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.