Abstract
This study proposed a CCD-based (charge-coupled device) optical computed tomography scanner (CT-s2) for 3D gel dosimetry. A parallel laser light was generated to pass through the gel sample using a diffuser and collimating lens. A CCD was used to capture projection image of gel sample at each step. An image reconstruction algorithm, filtered-back projection (FBP) technique was used to reconstruct the 3D image. Two better rotational steps are suggested as 1.0 degree and 1.5 degree for considering both of angular resolution and position deviation. The un-irradiated and irradiated BANG gel samples were scanned and reconstructed using FBP technique. Some artifacts were found in reconstructed images. Some discussions for artifacts were conducted and some solutions provided by previous researches to reduce these artifacts will be evaluated in the future work.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.