The applicability of a capillary zone electrophoresis–electrospray ionisation tandem mass spectrometric (CZE–ESI-MS–MS) method for the separation of nine fluoroquinolones was investigated. Method optimisation involved systematic trouble-shooting starting with the type and duration of capillary pre-washing and conditioning, the choice of both the CE run buffer, MS sheath liquid, CE run potential, ESI spray voltage, sheath gas flow-rate, MS capillary voltage and CE capillary and MS capillary temperatures. Another extremely important factor was found to be the degree to which the CE capillary protrudes into the ESI chamber as well as whether or not sheath gas and spray voltage are employed during the CE injection or not. The importance of the latter has, to our knowledge, not been addressed elsewhere. Nine fluoroquinolones have been separated and detected in a single run by this technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.