Abstract

A new calibration facility for measuring the thoron ( 220 Rn) activity concentration in air is currently being developed at the Physikalisch-Technische Bundesanstalt (PTB). The system consists of a reference chamber of well-defined volume and several electrodeposited 228 Th sources. In order to produce and maintain a homogeneous and constant activity concentration, a flow field is created inside the reference chamber and monitored. The 220 Rn activity concentration is measured with a multi-wire ionisation chamber (MWIC) operating in pulse mode and externally connected to the reference chamber. With this method thoron fields with activity concentrations ranging from 1 to 10 kBq/ m 3 are presently produced with an overall relative standard uncertainty of less than 5%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.