Abstract

In this paper, a biological protocol for endotoxin detection has been developed and optimized by quartz crystal microbalance (QCM). The parameters involved in the formation of the self-assembled monolayer (SAM) have been analyzed, and a study of the pH of the ligand buffer has been performed in order to find the best condition for the ligand immobilization and, in consequence, for the endotoxin detection. The detection limit obtained with the characterized biological protocol corresponds to 1.90 μg/ml. The effectiveness of the optimized biological protocol has been analyzed by cyclic voltammetry analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.