Abstract

A novel biosensor assay format for aflatoxin based on acetylcholinesterase (AChE) inhibition by aflatoxin B1 (AFB1) is proposed. The AChE was present in solution and an amperometric choline oxidase biosensor was used for monitoring its residual activity. To create the biosensor, the choline oxidase was immobilized by cross-linking onto screen-printed electrodes modified with Prussian Blue (PB) and these were used to detect the H2O2 at low potential (−0.05V versus a screen-printed internal silver pseudoreference electrode).For the development of the AFB1 assay, several parameters such as AChE and substrate concentration, the methanol effect, and pH were evaluated and optimized. The linear working range was assessed to be 10–60ppb. Concentrations as low as 2ppb, which correspond to the legal limit of AFB1 in food for humans, were detected after a pre-concentration step. The suitability of the method was evaluated using commercial olive oil samples. A recovery equal to 78±9% for 10ppb of AFB1 in olive oil samples was obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.