Abstract

In particle therapy, it is important to monitor the Bragg-peak position. It was simulated by GEANT4 Monte Carlo Simulation Code that the distribution of secondary generated gamma rays on the carbon beam therapy and the proton beam therapy. This simulation shows that gamma rays whose energy is 10 MeV or more are intensively generated at the Bragg-peak position. We are developing the system to monitor the Bragg-peak position which can measure pair production events occurred in the detector by gamma rays from irradiation points. The momentum direction of the gamma ray can be determined by measuring passing points and energy of e+ and e− generated by pair production. This system has 5 parts. The first is the conversion part. This part consists of several layers. Each layer is composed of a La-GPS ((Gd0.75La0.24Ce0.01)2Si2O7) scintillator plate and wavelength-shifting fibre (WLSF) sheets. The scintillator plate is sandwiched between sheets, where the directions of the sheets are in orthogonally x and y directions. In this part, gamma rays are converted to e+ e- pairs and the position where the conversion occured is determined. The second is the tracking part. This part consists of 2 layers of scintillating fibre tracker. Each layer has 6 scintillating fibre sheets for x, x’, u, u’, v, and v’. The third is the energy measurement part. It measures the energy of e+ and e− by scintillator array and Silicon Photomultipliers. The fourth is the veto counter for bremsstrahlung gamma rays from e+ and e-. The fifth is the beam monitor. By experiment, the number of photoelectrons of La-GPS with a WLSF (B-3(300)MJ, Kuraray) sheet and scintillating fibre (SCSF-78, Kuraray) when charged particle passed was measured as 9.7 and 7.6 respectively.

Highlights

  • I N particle therapy, cancer cells can be killed by matching the Bragg-peak position and the cancer position

  • We simulated on GEANT4 Monte Carlo Simulation Code that the distribution of secondary generated gamma rays on the carbon beam therapy and the proton beam therapy

  • The momentum direction of the gamma ray can be determined by measuring passing points and energy of e+ and e− generated by pair production

Read more

Summary

Introduction

I N particle therapy, cancer cells can be killed by matching the Bragg-peak position and the cancer position. In particle therapy, it is important to monitor the Bragg-peak position. To monitor the Bragg-peak position, positron emission tomography (PET) detectors [1], Compton cameras [2] have been studied. This simulation shows that gamma rays whose energy is 10 MeV or more are intensively generated at the Bragg-peak position. Gamma rays whose energy is around 0.511 MeV are widely

Results
Conclusion