Abstract

An enzymatic biosensor has been developed for the determination of selenate (SeO4 2- ), in which selenate reductase (SeR) is chemically attached to a gold disk electrode by lipoic acid N-hydroxysuccinimide ester as linker, allowing the catalytic reduction of the SeO4 2- to SeO3 2- . Modification of the gold electrode was characterized by X-ray photoelectron spectroscopy (XPS), time-of-flight secondary ion mass spectroscopy (ToF-SIMS), and electrochemistry. Cyclic voltammetry (CV) and differential pulse voltammetry (DPV) measurements were performed in different buffers for selenate determination. Under optimum conditions, the calibration curve was linear over the range 7.0-3900.0 μg L-1 with limits of detection and quantification of 4.97 and 15.56 μg L-1 , respectively. The possible interference of the relevant oxyanions SO4 2- , NO3 - , NO2 - , PO4 3- and AsO4 3- in the determination of SeO4 2- was studied. Finally, the proposed biosensor was used to determine SeO4 2- with recovery between 95.2 and 102.4 % in different real water samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.