Abstract

In this work, the development of a disposable electrochemical device (US$ 0.02 per electrode) using a 3D printed support (3Ds) of acrylonitrile butadiene styrene (ABS) insulating filament with a composite material (CM) based on graphite and nail polish, immobilized on the support surface, was described for the electrochemical determination of diclofenac (DCF). The device was compared to the commercial glassy carbon electrode (GCE) and showed superior electroanalytical performance with approximately 1.8-fold higher current density. Additionally, an amperometric method for DCF determination in tap water, synthetic urine, and pharmaceutical formulation samples with the proposed electrode, using a flow injection analysis (FIA-AD) system, was developed. The optimized method presented excellent detectability (LOD = 0.47µmol L-1), with excellent precision and accuracy (relative standard deviation < 5.6%) and percent recovery from spiked samples ranging from 89 to 106%. In addition, the sensor showed optimal analytical frequency with approximately 108 injections per hour, which demonstrates the potential of this system using the proposed disposable electrode for implementation in routine analysis and quality control with good selectivity and sensitivity.