Abstract
This paper introduces a newly constructed type-N coaxial microcalorimeter that will serve as a primary power standard in Korea Research Institute of Standards and Science. The detailed design scheme of the microcalorimeter is presented, and its characteristics are evaluated. The performance of an adiabatic line was simulated with a finite-element-method-based tool and predicted. The measured insertion loss of the assembled adiabatic line was less than 0.15 dB. Frequency and power sweep measurements were performed on a reference standard to find a substituted power level that minimized the dual-element error. We also propose evaluation standards that can be used for the measurement of radio frequency loss. The correction factor for the effective efficiency of the reference standard was experimentally evaluated. We also present an uncertainty evaluation process and the corrected effective efficiency of a type-N coaxial reference standard.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.