Abstract

We developed a 20[Formula: see text]MeV particle-induced X-ray emission (PIXE) analysis method using a medical cyclotron, which is conventionally used for positron emission tomography analysis performed in vacuo, during which the target sample is damaged. For non-destructive analysis and ease of switching between target samples, we developed a technique allowing 20[Formula: see text]MeV proton PIXE analysis to be performed at the atmospheric pressure. We filled the PIXE analysis chamber with helium and checked that the continuous background of the Compton tails of nuclear reaction [Formula: see text]-rays increased only minimally, and that the quasi-free electron bremsstrahlung (QFEB) did not increase at all, in the X-ray energy spectrum.