Abstract
Postural sway is a fall risk assessment tool traditionally measured using forceplates. However, it is expensive and lack portability to usage in clinical environments. In this study, a wearable low-cost MEMS inertial sensor was developed and validated for sway assessment. This study had two principal objectives (1) To develop and validate an Inertial Measurement Unit (IMU) for sway analysis (2) To assess the feasibility of IMU in detecting slight balance impairments due to reduced proprioception or induced subthreshold vibrations on the feet. A total of 10 healthy young adults (five males and five females) participated in this study. We found that sway velocity, Root Mean Square (RMS), and sway path length could successfully differentiate postural changes due to varying proprioceptive and sub-threshold vibration conditions. We conclude that new cheap wearable sensors can be designed for accurate clinical postural sway assessments and have potential to diagnose postural impairments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.