Abstract
A simple, accurate UV-VIS Spectrophotometric method has been developed using a dissolution media, Tween 20 (0.15%) as a solvent to determine the Triamcinolone. At predetermined λmax of 240 nm, method was proved linear in the range of 10-50 µg/ml with good correlation coefficient (R2= 0.998) and excellent recovery (between 98 and 101%). This method was successfully applied to the determination of Triamcinolone from dissolution media, 0.15% Tween 20 aqueous solution. This method can be employed for the dissolution study of Triamcinolone in injectable suspension.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Scientific Research in Science and Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.