A stability-indicating reversed phase high performance liquid chromatographic assay procedure has been developed and validated for etodolac. The liquid chromatographic separation was achieved isocratically on C18 Zorbax ODS using a mobile phase containing methanol: water: acetic acid (70:30:0.1%, v/v/v), at a flow rate 1 mL/min and UV detection at 254 nm. The method was linear over the concentration range of 2.4–16 µg/mL (r = 0.9999) with a limit of detection and quantitation of 0.03 and 0.10 µg/mL, respectively. Another method was applied for the determination of etodolac in the presence of its degraded products using the first derivative ratio spectrophotometry (DD1) at 293 nm using 1000 µg of degraded etodolac as a divisor. The method was linear over the concentration range of 10–50 µg/mL (r = 0.9999). Both methods have been found to have the required accuracy, selectivity, sensitivity, and precision to assay etodolac in bulk form and in pharmaceutical dosage form. Degradation products resulting from the stress studies did not interfere with the detection of etodolac, which indicates that the assays are stability-indicating assays.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.