Abstract

Objective: The objective of the present study was to develop and validate a novel stability indicating reverse phase-high performance liquid chromatography (RP-HPLC) method for determination of β-acetyldigoxin, an active pharmaceutical ingredient (API).Methods: The chromatographic separation was carried out on Agilent Technologies 1200 series HPLC system equipped with photo diode array detector and C-18 (4.6x250 mm, 5 µ) column. The mobile phase consisted of water: acetonitrile (65:35 v/v), delivered at a flow rate of 1.5 ml/min and eluents were monitored at 225 nm.Results: The retention time of β-acetyldigoxin was 9.2 min. The method was found to be linear (R2= 0.9995) in the range of 31.25-500 µg/ml. The accuracy studies showed the mean percent recovery of 101.02%. LOD and LOQ were observed to be 0.289 µg/ml and 0.965 µg/ml, respectively. The method was found to be robust and system suitability testing was also performed. Forced degradation analysis was carried out under acidic, alkaline, oxidative and photolytic stress conditions. Significant degradation was observed under tested conditions, except for oxidative condition. The method was able to separate all the degradation products within runtime of 20 min and was able to determine β-acetyldigoxin unequivocally in presence of degradation products.Conclusion: The novel, economic, rapid and simple method for analysis of β-acetyldigoxin is reported. The developed method is suitable for routine quality control and its determination as API, and in pharmaceutical formulations and stability study samples.

Highlights

  • Analytical grade β-acetyldigoxin was purchased from Clearsynth Labs, Mumbai

  • The method for analytical determination of β-acetyldigoxin was validated and tested for system suitability according to International Conference on Harmonization (ICH), United States Food and Drug Administration (FDA) and United States Pharmacopoeia (USP) guidelines [18,19,20]

  • The results complied with the acceptance criteria laid down under regulatory guidelines [20, 23] with % RSD values, tailing factor, asymmetry factor and height equivalent to theoretical plates being well below the acceptable limits and correlation coefficient, and number of theoretical plates were above the minimum acceptance criterion

Read more

Summary

Methods

The chromatographic separation was carried out on Agilent Technologies 1200 series HPLC system equipped with photo diode array detector and C-18 (4.6x250 mm, 5 μ) column. The mobile phase consisted of water: acetonitrile (65:35 v/v), delivered at a flow rate of 1.5 ml/min and eluents were monitored at 225 nm

Results
INTRODUCTION
MATERIALS AND METHODS
Method validation and system suitability testing
RESULTS AND DISCUSSION
CONCLUSION