A new simple, accurate, precise and selective stability-indicating high performance thin layer chromatographic (HPTLC) method has been developed and validated for the determination of Rizatiptan Benzoate in bulk and tablet formulation. Chromatographic separation was performed on aluminum plate precoated with Silica Gel 60 F254 using, Methanol: n-Propane: Triethylamine (3:5:2 v/v/v) as mobile phase followed by densitometric scanning at 280 nm. This system was found to give compact spot for Rizatiptan Benzoate (Rf value of Rf = 0.38 ± 0.02). The calibration curve was found to be linear between 800-2800 ng/band. The limit of detection and quantitation were found to be 2.24 ng/band and 6.79 ng/band respectively. The proposed method can be applicable for the estimation of Rizatiptan Benzoate during stability studies. Keywords: Rizatiptan Benzoate, High Performance Thin Layer Chromatography (HPTLC), Stability-Indicating Method

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.