Abstract

High-performance liquid-chromatographic (HPLC) methods were validated for determination of pravastatin sodium (PS), fluvastatin sodium (FVS), atorvastatin calcium (ATC), and rosuvastatin calcium (RC) in pharmaceuticals. Two stability-indicating HPLC methods were developed with a small change (10%) in the composition of the organic modifier in the mobile phase. The HPLC method for each statin was validated using isocratic elution. An RP-18 column was used with mobile phases consisting of methanol–water (60:40, v/v, for PS and RC and 70:30, v/v, for FVS and ATC). The pH of each mobile phase was adjusted to 3.0 with orthophosphoric acid, and the flow rate was 1.0 mL/min. Calibration plots showed correlation coefficients (r) > 0.999, which were calculated by the least square method. The detection limit (DL) and quantitation limit (QL) were 1.22 and 3.08 µg/mL for PS, 2.02 and 6.12 µg/mL for FVS, 0.44 and 1.34 µg/mL for ATC, and 1.55 and 4.70 µg/mL for RC. Intraday and interday relative standard deviations (RSDs) were <2.0%. The methods were applied successfully for quantitative determination of statins in pharmaceuticals.