Abstract

A convenient, simple, accurate, precise and reproducible Reverse Phase High Performance Liquid Chromatographic method was developed and validated for the estimation of candesartan cilexetil in bulk drug and tablet dosage form. The separation was achieved in less than 5 minutes using Dionex C18 column on 250 × 4.6 mm id, 5 μm particle size, mobile phase acetonitrile and phosphate buffer pH 4.0 in the ratio of 70:30 v/v at a flow rate of 1 mL/min. The effluent was monitored at 254 nm using diode array detector. Injecton volume applied was 10 μL The retention time of candesartan cilexetil was found to be 2.14 minutes and the standard calibration plot was found linear over the range of 10-200 ppm. The LOD and LOQ were found to be 5.61 ppm and 17 ppm respectively. The developed method was validated according to International Conference on Harmonization (ICH) guidelines for specificity, linearity, range, accuracy, precision, detection limit, quantitation limit, robustness and system suitability parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.